No, this is not possible.
Sisgeo calibrates each probe individually, but each probe has its own small different properties, that have a certain influence on the readings.
The international standard ISO 18674-3, describing the reference reading, states that:
"It is good practice to take an additional reference measurement with an independent probe that can be used as a backup".
The direction suggested by the standard is therefore clear: if you change the measurement probe, you must take a new reference measurement with the new probe.
However, with the KLION inclinometer data management software, Sisgeo offers a possibility to compare data from different probes. The method proposed by KLION can be applied by accepting 3 strong assumptions (see also the KLION software user manual at the following link, starting on page 43:https://www.sisgeo.com/uploads/manuali/KLION_SOFTWARE_EN_03_2022.pdf):
1 - there was no movement between the first measurement with the new probe and the last reading with the old probe, so they can be considered to be coincident.
2 - the international standard ISO 18674-3 considers measurements taken with different probes to be non-comparable.
3 - international publications suggest not using this approach: the correct procedure when starting measurements on an existing inclinometer tube with a new probe is to start a new data analysis and processing by acquiring a new reference measurement.
By accepting these assumptions, it will be possible to apply the 'new probe' function.
Comments
0 comments
Please sign in to leave a comment.